Beam Height | 10 |
Data Logger | 2000 measured values (Result, Date, Time, Sample ID, User ID) |
Dimensions (H x W x D) | 151 mm x 350 mm x 255 mm |
Display | 7″ TFT |
Display Resolution | WVGA (800 pix x 480 pix) |
Display Size | 7 inch (17.8 cm) |
Display Type | Colored touch-screen |
Enclosure Rating | IP40 |
Interfaces | USB type A (2), USB type B, Ethernet, RFID module |
Light Source | Gas-filled Tungsten (visible) |
Operating Conditions | 10 – 40 °C (50 – 104 °F), max. 80% relative humidity (non-condensing) |
Operating Mode | Transmittance (%), Absorbance and Concentration, Scanning |
Optical System | Reference beam, spectral |
Photometric Accuracy | 5 mAbs @ 0.0 – 0.5 Abs, 1 % at 0.50 – 2.0 Abs |
Photometric Linearity | < 0.5 % – 2 Abs |
Photometric Linearity 2 | ≤ 0.01 % at >2 Abs with neutral glass at 546 nm |
Photometric Measuring Range | ± 3.0 Abs (wavelength range 340 – 900 nm) |
Power Requirements (Voltage ) | 110 – 240 VAC |
Power Requirements (Hz) | 50/60 Hz |
Preprogrammed Methods | > 240 |
Sample Cell Compatibility Rectangular | 10, 50 mm, 1 inch; round: 13 mm, 16 mm, 1 inch |
Scanning Speed | > 8 nm/S (in steps of 1 nm) |
Source Lamp | Gas-filled Tungsten (visible) |
Specific Technology | RFID for easy method update, sample ID and Certificate of Analysis |
Spectral Bandwidth | 5 nm |
Stray Light | < 0.1% T at 340 nm with NaNO2 |
User Programs | 100 |
Wavelength Accuracy | ± 1.5 nm (wavelength range 340 – 900 nm) |
Wavelength Calibration | Automatic |
Wavelength Range | 320 – 1100 nm |
Wavelength Reproducibility | ± 0.1 |
Wavelength Resolution | 1 |
Wavelength Selection | Automatic, based on method selection |
Weight | 4.2 kg (9.3 lbs) |